Skip to content Skip to sidebar Skip to footer

Fume Hood CompactLine DCL 1200 PP

POL-EKO CompactLine Fume Hood DCL 1200 PP The DCL 1200 PP is a 1200 mm wide ducted fume hood engineered for handling aggressive chemicals in industrial and research laboratories. Built with a robust polypropylene structure and workspace, it provides maximum protection against corrosion and acid vapors. It features a motorized safety glass sash, integrated LED lighting, and an advanced microprocessor controller to monitor airflow velocity (face velocity). With its compact, aerodynamic design, it ensures the safe removal of hazardous fumes while optimizing laboratory space and operator safety.

Description

The CompactLine DCL 1200 PP is the ideal solution for laboratories performing heavy acid digestion, trace metal analysis, or any process involving concentrated mineral acids (like $HCl$, $H_{2}SO_{4}$, or $HF$). Unlike standard fume hoods that may use coated steel, the polypropylene (PP) construction of the DCL 1200 PP is inherently resistant to chemical attack, ensuring a long service life without the risk of rust or structural degradation.Advanced Safety and Airflow Management:Operator safety is the core focus of the CompactLine series:Airflow Monitoring: The unit is equipped with a Schneider airflow controller that monitors the face velocity in real-time.1 If the airflow drops below a safe threshold (e.g., $0.5 text{ m/s}$), an audible and visual alarm is triggered.Motorized Sash: The safety glass window is motorized, allowing for easy adjustment. It includes a “stop” mechanism to prevent the sash from being opened beyond the safe working height.Aerodynamic Design: The hood’s interior is shaped to minimize turbulence and “dead zones,” ensuring that heavy vapors are efficiently pulled toward the exhaust baffles.Ergonomics and Integrated Utilities:Despite its compact “tabletop” style footprint, the DCL 1200 PP is fully featured:Illumination: Integrated LED lighting provides a bright, shadow-free workspace (>800 lux).2Worktop Options: While the interior is PP, the worktop can be customized with materials such as solid ceramics, epoxy resin, or technical stoneware depending on the specific heat and chemical requirements.Service Modules: The side pillars are equipped with electrical sockets (IP44 protected) and can be fitted with remote-controlled valves for water, gas, or vacuum.Primary Applications:Analytical Chemistry: Acid digestion and sample preparation for ICP-MS/OES.Battery Research: Handling of electrolyte components and corrosive salts.Semiconductor Lab: Etching processes using aggressive acids.General Research: Any application requiring a 1200 mm wide workspace with high chemical resistance.

Datasheet

Specifications

Fume Hood CompactLine DCL 1200 PP Specs

The DCL 1200 PP is a high-resistance laboratory fume hood from the POL-EKO CompactLine series. The “PP” designation indicates that the internal working chamber is finished with polypropylene, providing excellent protection against aggressive inorganic acids, bases, and salts.

 
  • Internal Chamber Material: Polypropylene (PP)

     
  • Worktop Material: Monolithic ceramics with a raised marine edge (Standard)

     
  • External Dimensions (W x H x D): 1280 x 2325–2600 x 960 mm

     
  • Working Space Dimensions (W x H x D): 1150 x 1220 x 635 mm

     
  • Front Window: Toughened glass, manually operated with a counterweight; opens upwards with a lock at 500 mm (max opening 810 mm), including additional horizontal sliding glass panels.

     
  • Airflow Management: Double back-wall system with upper and lower extraction slots to remove vapors of various densities without turbulence.

     
  • Monitoring System: Schneider FM 550-A-0-E controller with an air velocity sensor (compliant with EN 14175).

     
  • Lighting: Energy-saving linear LED lighting, separated from the chamber by a glass pane.

     
  • Standard Fittings:

    • 2 x Water spouts with front-panel valves (BROEN)

       
    • 1 x Ceramic sink (280 x 80 mm)

       
    • 2 x Electrical sockets (230V, 50Hz)

       
  • Safety Features: * Safety flap in the ceiling for decompression.

     
    • Protection system to prevent uncontrolled window drops.

       
  • Airflow Parameters: 600 – 950 m³/h (recommended)

     
  • Exit Air Duct Diameter: Ø 200 mm

     
  • Nominal Power: 46 W

     
  • Voltage: 230V, 50/60 Hz

     
  • Warranty: 24 months

     

Key Material Advantage

While the DCL 1200 CR uses ceramic tiles for general chemical and high-heat resistance, the DCL 1200 PP is specifically chosen for environments where corrosion from acids is the primary concern. Polypropylene is a non-metallic, non-porous material that prevents the degradation often seen in galvanized or ceramic-lined hoods when exposed to constant acid fuming.