Skip to content Skip to sidebar Skip to footer

Climatic Chamber KKP 240

POL-EKO KKS 1450 Climatic Chamber The KKS 1450 is a high-capacity, 1564-liter double-door climatic chamber engineered for the most demanding large-scale stability and environmental tests. Featuring a silent, vibration-free steam humidification system, it provides a stable temperature range of 0?C to +60?C and humidity from 10% to 90% RH. Managed by the Smart PRO 7-inch touchscreen controller, the KKS 1450 offers advanced multi-segment programming, Wi-Fi connectivity, and full GLP-compliant data logging. It is the premier choice for high-throughput ICH pharmaceutical trials, large-batch seed germination, and insect breeding.

SKU: KKP 240 Categories: , Brand: Product ID: 21635

Description

The POL-EKO KKS 1450 represents the pinnacle of capacity and precision in climatic simulation. With its massive 1564-liter internal volume and double-door architecture, this unit is designed for laboratories and industrial facilities that must manage thousands of samples or oversized components simultaneously. By employing steam humidification technology, the KKS 1450 avoids the high-frequency vibrations and noise of ultrasonic systems, creating a stable, “silent” environment that is essential for delicate biological applications, such as the mass breeding of Drosophila melanogaster and long-term plant growth studies.

Intelligent Control & Global Connectivity: Equipped with the top-tier Smart PRO controller, the KKS 1450 offers a high-resolution 7-inch color touchscreen interface. Users can program up to 40 custom profiles with 100 segments each, simulating complex day/night cycles or multi-week stability ramps.

Data Integrity: Integrated Wi-Fi, LAN, and USB ports allow for remote monitoring and effortless data export. The system includes an internal memory for 10,000 records and is fully compatible with LabDesk software for GLP/GMP-compliant reporting.

Unmatched Safety: Standard features include Class 3.3 over-temperature protection (DIN 12880), “Open Door” alarms, and the signature “Alarm Bar” for immediate visual status recognition across the laboratory.

Industrial-Grade Construction: The KKS 1450 is built for longevity with an interior of acid-proof stainless steel (DIN 1.4301). Its double-door system?comprising two insulated solid outer doors and two internal glass doors?allows researchers to access or inspect specific sections of the chamber without a total loss of the internal microclimate. The unit is mounted on heavy-duty lockable castors for mobility and includes a standard 30mm access port for external validation sensors.

Primary Applications:

Pharmaceuticals: High-throughput stability testing according to ICH Q1A (R2) guidelines.

Agriculture & Botany: Large-scale phytotron research, seed germination, and plant physiology (available with FIT options).

Industrial Testing: Accelerated aging and stress testing of large electronics, plastics, and packaging.

Life Sciences: Vibration-free mass breeding of insects and microorganisms.

Specifications

General Description

  • Model: KKP 240 climatic chamber with forced air convection, humidity control and Peltier cooling system (quiet, vibration-free).

  • Capacity: ~245 L chamber / working capacity 245 L.

  • Controller: Microprocessor PID with 7″ full-colour touchscreen display.

  • Temperature range (no humidity): 0…+70 °C (minimum ~20 °C below ambient).

  • Temperature range (with humidity): +5…+70 °C.

  • Temperature resolution: 0.1 °C.

  • Temperature spatial variation: ±0.3 °C at 40 °C; temporal fluctuation: ±0.1 °C.

  • Relative humidity range: 10 – 90 % RH.

  • Humidity resolution: 0.1 % (variability: spatial <±1 % RH, temporal ±0.3 % at 40 °C, 75 % RH; spatial ±0.8 %, temporal ±0.2 % at 25 °C, 60 % RH).

  • Recovery time: ~10 min to set values after a 30 s door open at 40 °C, 75 % RH.

  • Protection class: Over-temperature protection class 3.3 to DIN 12880.

Chamber Construction

  • Doors: One external solid door + one internal glass door.

  • Interior: Acid-proof stainless steel to DIN 1.4301; housing: powder coated sheet (Smart PRO version).

  • Access port: Ø 30 mm for external sensors.

  • Shelves: 3 standard, up to 10 optional; max shelf load: 25 kg; max unit load: 90 kg.

  • Water tank: External 6 L container for deionised water for humidity generation.

Dimensions & Weight

  • Overall dimensions: Width ~820 mm, Width with water tank ~960 mm, Height ~1140 mm, Depth ~840-940 mm (depending on configuration).

  • Internal dims: Width ~600 mm, Height ~800 mm, Depth ~500 mm.

  • Weight: ~117 kg.

Electrical

  • Voltage: 230 V, 50/60 Hz.

  • Nominal power: ~2300 W.

Features & Safety

  • Forced air circulation for uniform conditions.

  • Open-door alarm and multi-segment time/temperature programming.

  • LAN & USB ports for data download and connectivity.

  • Wi-Fi communication and LabDesk software support (optional).

Typical Applications
Suitable for ICH stability testing, environmental conditioning, plant cultivation, material testing, insect breeding and other research requiring controlled temperature and humidity.

Datasheet