Skip to content Skip to sidebar Skip to footer

Walk-In Fume Hood CompactLine DCL 1200

The Walk-In Fume Hood CompactLine DCL 1200 is designed for safe handling of large laboratory equipment, offering reliable airflow control, durable construction, and certified laboratory safety performance.

Description

The Walk-In Fume Hood CompactLine DCL 1200 by POL-EKO is a professional laboratory ventilation solution engineered for applications requiring accommodation of large or floor-standing equipment. Its walk-in design ensures unrestricted access while maintaining effective containment of hazardous fumes and vapors.

Constructed from robust galvanized steel, the CompactLine DCL 1200 features a tempered glass sash, efficient airflow extraction system, and certified monitoring compliant with laboratory safety standards. The unit provides reliable protection for users, ergonomic operation, and long-term durability, making it ideal for research, chemical, industrial, and educational laboratories requiring dependable ventilation for demanding workflows.

Specifications

Walk-In Fume Hood CompactLine DCL 1200 Specifications

The Walk-In (Floor-Mounted) CompactLine DCL 1200 is designed for laboratory procedures requiring large-scale apparatus, tall distillation columns, or heavy equipment that must be rolled directly into the workspace on a trolley. By eliminating the standard worktop and base cabinet, it provides a floor-level entry and a high internal working chamber.

Technical Parameters

  • Recommended Airflow: 900 – 1200 m3/h

  • Air Velocity: 0.3 – 0.5 m/s

  • Nominal Power: 60 W

  • Voltage: 230 VAC 50Hz

  • Air Outlet Diameter: 200 mm or 250 mm

  • Lighting Intensity: > 800 LUX (External LED)

  • Noise Level: < 56 dB

Physical and Material Data

  • Overall Dimensions (W x H x D): 1200 x 2350 x 900 mm

  • Working Space Dimensions (W x H x D): 1134 x 2100 x 730 mm

  • Body Construction: Galvanized steel with chemically resistant epoxy powder coating

  • Internal Liner: Epoxy-coated galvanized steel (Optional: Polypropylene or Phenolic Resin)

  • Front Window: Two-part telescopic toughened glass sash with manual counterweight

  • Floor: Laboratory floor or optional low-profile ceramic/stainless steel floor tray

  • Weight: Approx. 580 kg

Functional and Safety Features

  • Triple-Level Extraction: The walk-in version features a specialized rear baffle system with extraction slots at the top, middle, and floor levels. This ensures that heavy vapors settling near the floor are effectively removed along with lighter gases.

  • Telescopic Sash System: Features a vertically sliding window consisting of two independent panes. This allows for a very high opening for equipment setup while maintaining a safe working aperture during operation.

  • Floor-Level Access: The design allows for the easy introduction of mobile laboratory tables, large reactors, or drums without the need for lifting over a sill.

  • Airflow Monitoring: Equipped with the Schneider FM 550 controller which provides continuous monitoring of air velocity and triggers visual/acoustic alarms if conditions fall below EN 14175 safety levels.

  • Decompression System: A safety relief flap is integrated into the ceiling to safely vent pressure in the event of a rapid reaction or explosion inside the chamber.

  • External Utilities: Service valves and electrical sockets (IP44) are mounted on the side pillars outside the contaminated zone, providing easy control of gas, water, and power.

  • Anti-Drop Protection: The sash is equipped with a safety mechanism that prevents the window from falling in the event of a drive cable failure.